-40%

Dry Block Temperature Calibrator Range 300-1200C 5.7''TFTLCD Touch Screen AC220V

$ 1425.6

Availability: 21 in stock
  • MPN: ET3801-1200
  • Condition: New
  • Brand: Unbranded
  • Country/Region of Manufacture: China

    Description

    Note: can work with AC 220V only, not for AC 110V power supply.
    ET3801-1200
    This product is easy to carry, easy to use, fast and reliable temperature efficiency, widely used in machinery, chemical industry, food, medicine and other industries. The problem of slow temperature rise and slow temperature rise is common in the existing dry test furnace in China, which leads to a long time for users to calibrate each time. The latest generation of dry well furnace of our company adopts the most advanced heating principle design in the world. It has the characteristics of fast temperature rise, fast temperature rise and fast temperature drop, which greatly improves the existing calibration efficiency.
    Through reliable high precision temperature control circuit and high precision sensor, the precision is higher than that of other domestic manufacturers, and the technology has reached international standards. The first worldwide touching operation, simple and fast to use.
    Main features:
    Small
    size,
    light
    weight,
    easy
    to
    carry;
    Various
    types
    of
    insertion
    tube,
    can
    meet
    the
    different
    size,
    quantity
    of
    sensor
    testing
    and
    calibration.
    And
    can
    be
    customized
    according
    to
    the
    special
    needs
    of
    users;
    Horizontal
    temperature
    and
    vertical
    temperature
    field;
    Check
    the
    depth
    of
    insertion
    with
    the
    industry
    leader;
    5
    .7
    inch
    TFTLCD
    touch
    screen,
    16
    bit
    true
    color
    RGB
    display,
    full
    touch
    operation,
    intuitive
    use;
    Fast cooling,
    easy
    to
    set,
    good
    temperature
    stability;
    Hot and
    fast replacement;
    With
    load
    short
    circuit,
    load
    switch,
    sensor protection
    and
    other functions.
    General technical specifications
    Language: Chinese, English;
    Communication interface: USB Device (optional).
    The ambient temperature range: 0~50℃(32-122℉);
    Ambient humidity range: 0%-90% (non condensing);
    model
    ET3801-150A
    ET3801-150B
    ET3801-650
    ET3801-1200
    temperature range
    -20℃~150℃
    -30℃~150℃
    50℃~650℃
    300℃~1200℃
    heating rate
    30~100:20min
    30~150:40min
    30~100:20min
    30~150:40min
    30~300:7min
    30-400:12min
    30~650:25min
    30~1200:75min
    Display accuracy
    ≤±0.1℃
    ≤±0.1℃
    Below 400℃

    ≤±0.35℃
    400-650℃

    ≤±0.5℃
    ≤±1.2℃
    Insertion depth
    160mm
    150mm
    135mm
    Average heat block diameter
    36mm
    32mm
    32mm
    39mm
    Temperature field stability
    ≤±0.02℃
    ≤±0.02℃
    ≤±0.05℃
    ≤±0.2℃
    Horizontal temperature field
    ≤±0.05℃
    ≤±0.05℃
    ≤±0.05℃
    ≤±0.25℃
    Vertical temperature field
    The deviation in the range of 50mm calculated from the bottom of the hole of the soaking block is 1 degree
    Temperature unit
    ℃ or ℉
    Temp Accuracy
    0.1% or 0.2%
    Display resolution
    0.01℃
    0.01℃
    0.01℃
    0.01℃
    maximum power
    Negative temperature type: 500W, medium temperature type: 800W, high temperature type: 3000W
    weight(Net weight)
    Negative temperature type: 13kg, medium temperature type: 11kg, high temperature type: 11kg
    weight (Containing packaging)
    Negative temperature type: 23kg, medium temperature type: 18kg, high temperature type:20kg. Packaging includes aluminium box and transport wooden box
    Outline size
    Negative temperature type and high temperature type: 310*190*340mm, medium temperature type: 250*150*310mm
    Use environment
    Ambient temperature 0-50℃, relative humidity less than 95% (no condensation)
    Power supply
    220VAC +- 10%, 45 - 60Hz
    Optional function
    Verification Data Recording and Exporting (ET3801 only optional), up to 250 sensors, 2500 records,USB Device Communication Interface
    Other temperature ranges should be specified when ordering. The negative temperature type of -20℃-150℃ has 4 holes,are φ6/ 8/ 10/ 12 mm .The negative temperature type of     -30 ~150 has 4 holes, are 6/ 8/ 8/1
    0
    mm. The medium temperature type of  has 2 kinds (2 choices 1    at the time of ordering), one is 3 holes of 8/ 10/ 12mm, the other is 4 holes of 6mm*2, 8mm*2. The high temperature     type of 300 -1200 has 4 holes, are 6/ 8/ 10/ 12mm.    Special specification aperture can be customized and should be specified when ordering.